ARTICLE

Ensure compliance

25 January 2013

The new T3 Gas Tester ensures Crowcon's Tetra:3 (T3) portable, multi-gas detector complies with relevant national/regional safety regulations and company requirements. The device feeds the T3's sensors a known conce

The new T3 Gas Tester ensures Crowcon's Tetra:3 (T3) portable, multi-gas detector complies with relevant national/regional safety regulations and company requirements.

The device feeds the T3's sensors a known concentration of gas, a process known as 'bumping', to ensure the sensors are working correctly. When a T3 is inserted into the Gas Tester, the T3 automatically switches into 'bump' mode and either passes or fails. Once the test is complete the T3 is removed and the bump result is then accepted by the user and logged automatically in the T3. When the test has passed, the user is safe in the knowledge their detector is in good working order; if it fails the bump, it informs the user that the instrument needs to be re-calibrated or serviced. All this is done without a power supply.

The Gas Tester can bump for flammable, oxygen, carbon monoxide, hydrogen sulphide and carbon dioxide sensors. It is capable of testing up to four gases within 60sec.
 
OTHER ARTICLES IN THIS SECTION
FEATURED SUPPLIERS
 
 